Eutech DO 2700 Dissolved Oxygen Meter (Model ECDO270042)

EUTECH
SKU:
EUT-MTP-ECDO270042
₹225,000.00
(No reviews yet)
Current Stock:
Minimum Purchase:
1 unit
Adding to cart… The item has been added
Product Description
 
The Eutech DO 2700 is a laboratory-grade benchtop meter engineered for optimal performance and versatility in water quality analysis. It is specifically designed to measure Dissolved Oxygen (DO), Biochemical Oxygen Demand (BOD), and Temperature. The intuitive meter offers advanced setup options for extensive user customisation and is ideal for routine BOD analysis and various other Dissolved Oxygen applications in the laboratory.
The meter is part of the Eutech 2700 Series of bench meters, known for delivering accurate and reliable results in laboratory settings.
 
Advanced Measurement and Compensation:
 
The DO 2700 ensures accurate readings across varying environmental conditions.
Versatile DO Measurement: It measures Dissolved Oxygen in units of % saturation, ppm, and mg/L, achieving an accuracy of ±0.5% full scale.
Enhanced Calibration: The system allows for automatic calibration at 100% saturation and an independent 0% point, which provides greater measurement sensitivity, particularly useful during low oxygen level testing.
Automatic Compensation: Crucial for accurate results in diverse samples, the meter features automatic compensation for Temperature, Salinity, and Barometric Pressure using a built-in sensor.
Data Management and GLP Compliance
The Eutech DO 2700 is equipped with several features designed to simplify data logging, transfer, and regulatory compliance.
Extended Memory: It features a non-volatile internal memory capable of storing up to 500 data points. Each stored data set is time and date-stamped for compliance with GLP (Good Laboratory Practice) standards.
Data Transfer: Data transfer to a PC or printer is simple using the bi-directional RS232 serial communication port.
Continuous Monitoring: An auto-logging function allows for convenient continuous monitoring of measurements.
Alarms and Security: The meter includes a Cal-due alarm to remind users when calibration is necessary, and a Limit alarm which alerts users if readings fall outside a pre-set range. Setup and calibration functions can be secured via password protection.
Design and Components
The meter is designed with convenience and durability in mind for laboratory use.
Self-Stirring Probe: The meter comes standard with a Self-Stirring Polarographic Dissolved Oxygen/BOD/ATC Electrode (EC620SSP), making it immediately suitable for BOD testing procedures.
Display: It features a comprehensive backlit LCD display, allowing simultaneous viewing of the main measurement reading along with the temperature.
Physical Design: The unit has a compact footprint, helping to conserve valuable bench space. It features a splashproof housing and keypad, offering protection against incidental laboratory spills.
Warranty: The Eutech meters in the 2700 series typically come with a 3-year warranty against manufacturing defects.
 
Typical Applications:
 
The Eutech DO 2700 is perfectly suited for professionals involved in:
Environmental studies
Wastewater and water treatment
Ecological studies
Educational institutions
 
Technical Specifications (Model ECDO270042):
 
Parameter
Range
Resolution
Accuracy
Dissolved Oxygen (DO)
0.00 to 50.00 mg/L (ppm)
0.01 mg/L
±0.5 % full scale
% Saturation Of Oxygen
0 to 600.0 %
0.1 %
±0.5 % full scale
Salinity Correction
0.0 to 50.0 ppt
0.1 ppt
-
Barometric Pressure Correction
450 to 825 mmHg (Automatic correction with built-in sensor)
1 mmHg
-
Temperature
0.0 to 50.0 °C / 32.0 to 122.0 °F
0.1 °C / 0.1 °F
±0.3 °C / ±0.5 °F
Calibration Points
2 points (100% and 0%)
-
-
Memory
500 data sets (GLP Compliant)
-
-
Output
Bi-directional RS232
-
-
Power Requirements
9 V DC adapter, 1.3 A (100/240 VAC, SMPS)
-
-
Meter Dimensions (L x W x H)
17.5 x 15.5 x 6.9 cm
-
-
Meter Weight
650 g
-
-