Eutech pH 2700 Benchtop pH, ORP and Temperature Meter
The Eutech pH 2700 is a laboratory bench meter for pH, oxidation-reduction potential (ORP) and temperature measurement. Its selectable pH resolution reaches 0.001 pH, with a manufacturer-specified accuracy of ±0.002 pH + 1 LSD. The ECPH270042GS configuration includes a refillable glass pH electrode and a separate automatic temperature compensation probe.
Calibration and Electrode Performance
Choose up to six preset calibration points or five custom buffers to match your measurement procedure. Automatic buffer recognition simplifies calibration with the selected buffer set, while slope and offset displays help assess electrode performance. Custom calibration buffers must be at least 1.0 pH unit apart.
Measurement Records for Laboratory Work
The meter stores up to 500 date- and time-stamped data points in non-volatile memory. Automatic logging and bidirectional RS232 communication support recording and transfer to a compatible computer or printer. The backlit display presents the measured value with temperature, electrode status and calibration information.
Select the Correct Electrode and Kit
This listing is for manufacturer kit ECPH270042GS, part number 01X543902. It includes the ECFG7370101B glass-body, double-junction refillable pH electrode, PH5TEMB01P temperature probe and 60 mL electrode refill solution. ORP measurement requires a suitable ORP electrode, which is not listed in this kit. Select an electrode appropriate for your sample and operating conditions.
SciSpectrum Lab Essentials supplies Eutech instruments as a reseller. When requesting a quotation, quote Sci SKU EUT-PH-2700 and kit ECPH270042GS, along with your sample type, expected pH range and any additional electrode requirements.
Refer to the Manufacturer Catalogue, SciSpectrum Specification Sheet and Manufacturer User Manual in Downloads for specifications, calibration procedures and maintenance guidance.
pH 2700 Technical Specifications
| Measurement parameters | pH, ORP / relative mV and temperature |
| pH range | -2.000 to 20.000 pH |
| pH resolution | 0.1 / 0.01 / 0.001 pH selectable |
| pH accuracy | ±0.002 pH + 1 LSD (least significant digit) |
| pH calibration | Up to 6 preset or 5 custom points; custom buffers at least 1.0 pH unit apart |
| Buffer sets | USA, NIST, DIN, User 1, User 2 and Custom |
| ORP / relative mV range | ±2000.0 mV |
| ORP resolution / accuracy | 0.1 mV / ±0.2 mV |
| Temperature range (meter) | 0.0 to 100.0 °C / 32.0 to 212.0 °F |
| Temperature resolution | 0.1 °C / 0.1 °F |
| Temperature accuracy | ±0.3 °C / ±0.5 °F |
| Temperature compensation | Automatic or manual (ATC / MTC) |
| Memory | 500 data points; non-volatile; date/time stamped |
| Communication | Bidirectional RS232; compatible PC or printer output |
| Display | Backlit display with measurement and electrode/calibration information |
| Meter dimensions / weight | 17.5 × 15.5 × 6.9 cm (L × W × H) / 650 g |
| Manufacturer kit / part number | ECPH270042GS / 01X543902 |
| Applications | Educational laboratories; ecological studies; environmental studies; chemical laboratories; titration measurements; quality assurance testing; food-processing laboratory measurements |
The temperature range above is the meter specification. Electrode compatibility and sample conditions should be checked separately.
Features and Practical Benefits
- 0.001 pH resolution: Select the display resolution required by your laboratory procedure.
- Multi-point calibration: Use up to six preset points or five custom buffers.
- 500-point memory: Retain date- and time-stamped measurements for review.
- RS232 data transfer: Output readings to a compatible computer or printer.
- Electrode diagnostics: View pH slope and offset information when assessing electrode condition.
- Calibration controls: Calibration-due reminders and password protection support consistent setup.
- Automatic logging: Record readings at configured intervals for monitoring.
- Backlit bench display: Review measured values and calibration information on one screen.
- Flexible electrode holder: Position the holder on either side of the meter to suit your bench.
ECPH270042GS Manufacturer Kit Contents
- Eutech pH 2700 meter
- ECFG7370101B glass-body, double-junction Ag/AgCl refillable pH electrode
- PH5TEMB01P stainless steel ATC probe
- 60 mL pH electrode refill solution
- Integral electrode holder
- 30X427301 RS232 cable
- 100/240 VAC power adapter
A separate ORP electrode and any additional calibration, storage or cleaning solutions should be specified according to your requirements. Consult the operating manual for calibration, electrode care and connection procedures.
pH 2700 Frequently Asked Questions
What does the Eutech pH 2700 measure?
It measures pH, ORP / relative mV and temperature with suitable electrodes. It does not provide the conductivity, dissolved oxygen or ion-concentration functions found on other 2700 Series models.
What is the pH measurement accuracy?
The manufacturer specifies ±0.002 pH + 1 LSD, with selectable resolution of 0.1, 0.01 or 0.001 pH over a -2.000 to 20.000 pH range. Accuracy and display resolution are different specifications.
Which electrode is included in ECPH270042GS?
The kit includes the ECFG7370101B refillable glass-body, double-junction pH electrode, a PH5TEMB01P ATC probe and 60 mL refill solution, together with the listed meter accessories.
Is an ORP electrode included?
An ORP electrode is not listed in ECPH270042GS. Request a suitable ORP electrode separately if you need redox measurements.
How many calibration points are supported?
The pH 2700 supports up to six preset points or five custom points. Custom buffers must be at least 1.0 pH unit apart. Follow the manual for buffer-set selection and calibration steps.
Can measurements be saved and transferred to a computer?
Yes. Non-volatile memory holds up to 500 date- and time-stamped data points. Bidirectional RS232 supports transfer to a compatible computer or printer; check connection and software requirements before setup.
Does the meter support automatic temperature compensation?
Yes. Automatic and manual temperature compensation are supported, and this kit includes a separate ATC probe. Temperature compensation does not remove the need to follow your sample measurement procedure.
Where can I find operating and maintenance instructions?
Open Downloads and select Manufacturer User Manual. The official 2700 Series manual includes pH 2700 instructions for setup, calibration, measurement, data handling and troubleshooting.
Product overview
Key features
Applications
Specifications
Applications
Features & benefits
Downloads
Frequently asked questions
Need help selecting a Benchtop pH Meter?
Tell us your application and our specialists will recommend the right model.