Formazin vs Polymer Turbidity Standards for Calibration and Verification
SciSpectrum Lab Essentials
Formazin vs Polymer Turbidity Standards for Calibration and Verification
You have two vials marked 100 NTU and an instrument waiting for its next calibration point. Before inserting either vial, read the kit number. The number on the vial identifies its assigned turbidity. The instrument manual and your method tell you whether that particular suspension belongs in the calibration.
This distinction matters when a lab runs several meters, replaces an expired kit or switches from bottled standards to sealed vials. Write down the exact meter, method, standard material and required points before you ask for a replacement.
Identify the material before comparing kits
Formazin is itself a polymer suspension. In supplier catalogues, “polymer standard” often means a different suspension based on synthetic microspheres, such as AMCO-AEPA-1. Stabilised formazin, including Hach StablCal, is a formazin formulation. US EPA Method 180.1 discusses both conventional formazin and the named AMCO-AEPA-1 standard. For your work, use the requirements of your applicable method and instrument. US EPA Method 180.1, sections 2 and 7

Match the kit to the meter and the calibration curve
Turbidity is an optical measurement. The particles, light source and detector arrangement affect the scattered-light response. Match the standard to the exact model and method, including the reporting unit and selected mode. For sample results, retain the method and unit together when comparing NTU and FNU readings from different instruments. US EPA Method 180.1, sections 2 and 7 · Thermo Scientific AQ4500 manual, vial handling and measurement
| Instrument and kit | What the manufacturer lists | What to confirm before ordering |
|---|---|---|
| Hach 2100Q, StablCal kit 2971210 | 20, 100 and 800 NTU standards with a 10 NTU verification standard. Hach 2100Q StablCal kit 2971210 | Selected curve, required points and bottle or sealed vial format. |
| Hanna HI98703 or HI93414, kit HI98703-11 | AMCO-AEPA-1 cuvettes with expected readings of <0.10, 15.0, 100 and 750 NTU. Hanna HI98703-11 turbidity standards | Exact instrument, required points and certificate values for the supplied lot. |

For a replacement order, put the kit number, standard type, points and intended use on the same line. A quote that says only “100 NTU standard” leaves those decisions open. Ask for the certificate of analysis and confirm that it covers the supplied lot.
Handle each vial according to its instructions
Read the instructions for the actual pack before mixing. A sealed suspension, a bottled standard and a solid check vial can require different handling. Where the instructions specify resuspension, use the prescribed motion and allow bubbles to clear as directed. Leave sealed vials sealed. If your procedure calls for preparing a working dilution from stock formazin, use the specified water, preparation steps and holding time. Follow the applicable safety data sheet for the material. US EPA Method 180.1, sections 2 and 7 · Hach guidance on Gelex verification standards
Check the glass before blaming the standard. Fingerprints, condensation, scratches and particles on the vial can affect the optical reading. Wipe the outside with a clean lint-free cloth and handle the vial away from the light path. Use the specified fill and orientation. If the manual calls for silicone oil, follow its application instructions. The Thermo Scientific AQ4500 manual gives model-specific guidance on vial handling and indexing. Thermo Scientific AQ4500 manual, vial handling and measurement
Record the vial ID and orientation with an unexpected result. If rotating a vial changes the reading, check the glass and indexing under the manual before using the result to adjust the instrument.
Verify the calibration before accepting sample results
Calibration sets the instrument response using the prescribed standards. Verification checks the response against an assigned value or established baseline. The Hach 2100Q kit lists a dedicated 10 NTU verification standard. Hach describes Gelex as a check material used with an instrument-specific baseline established after calibration. Follow the instructions for the actual check material. Hach 2100Q StablCal kit 2971210 · Hach guidance on Gelex verification standards

Suppose a check standard has an assigned value of 10.0 NTU and the meter reads 10.6 NTU. The difference is +0.6 NTU, or +6% relative to that assigned value. This is an arithmetic example. Compare the unrounded result where available with your method’s criterion, using its required calculation and rounding. Record the decision and the criterion together.
If a check falls outside the criterion, retain the original reading. Confirm the vial, lot, expiry, preparation, orientation and instrument mode. Repeat only as the procedure allows and keep the repeat with the original result. If the check still fails, follow your laboratory’s procedure for instrument status and review of affected sample results.
Keep a record the next analyst can use
Write down the calibration and verification materials separately. This helps the next analyst distinguish a new calibration from a repeat check and identify whether a replacement kit changed the vial format, standard material or assigned value.
| Record field | Evidence to retain |
|---|---|
| Instrument and method | Model, serial number, method, mode and reporting unit |
| Calibration material | Kit or part number, material, calibration points, lot and expiry |
| Vial preparation | Pack format, mixing or settling step, condition and orientation |
| Calibration record | Date, analyst, selected curve and stored calibration result |
| Verification material | Check ID, lot and assigned value or documented baseline |
| Observed check | Original reading, difference, acceptance criterion and decision |
| Follow-up | Investigation, repeat result, reviewer and instrument status |
Questions from the bench
Can I substitute a polymer vial with the same NTU value?
Is stabilised formazin the same as a microsphere standard?
Can the verification vial be used to calibrate?
What if a check changes when I rotate the vial?
For the measurement principles and routine water-testing workflow, see Turbidity in Water.
Identify the kit before you order
SciSpectrum Lab Essentials is a reseller of laboratory instruments and accessories. Send your turbidimeter model, current kit number, required calibration points and check standard. Include the vial format if your method specifies one.
Request a compatible standard kitTechnical references
- US EPA Method 180.1, sections 2 and 7
- Hach 2100Q StablCal kit 2971210
- Hanna HI98703-11 turbidity standards
- Hach guidance on Gelex verification standards
- Thermo Scientific AQ4500 manual, vial handling and measurement