Formazin vs Polymer Turbidity Standards for Calibration and Verification

SciSpectrum Lab Essentials
Two capped turbidity cuvettes beside a benchtop turbidimeter and the SciSpectrum logo.
Turbidity Measurement

Formazin vs Polymer Turbidity Standards for Calibration and Verification

SciSpectrum Lab Essentials6 min readWater TestingBench Guide

You have two vials marked 100 NTU and an instrument waiting for its next calibration point. Before inserting either vial, read the kit number. The number on the vial identifies its assigned turbidity. The instrument manual and your method tell you whether that particular suspension belongs in the calibration.

This distinction matters when a lab runs several meters, replaces an expired kit or switches from bottled standards to sealed vials. Write down the exact meter, method, standard material and required points before you ask for a replacement.

Identify the material before comparing kits

Formazin is itself a polymer suspension. In supplier catalogues, “polymer standard” often means a different suspension based on synthetic microspheres, such as AMCO-AEPA-1. Stabilised formazin, including Hach StablCal, is a formazin formulation. US EPA Method 180.1 discusses both conventional formazin and the named AMCO-AEPA-1 standard. For your work, use the requirements of your applicable method and instrument. US EPA Method 180.1, sections 2 and 7

Conventional formazin, stabilised formazin and synthetic microsphere standards compared by how the laboratory selects them.
Three types of turbidity standard Formazin is itself a polymer. In this guide, polymer standards refers to products based on synthetic microspheres.

Match the kit to the meter and the calibration curve

Turbidity is an optical measurement. The particles, light source and detector arrangement affect the scattered-light response. Match the standard to the exact model and method, including the reporting unit and selected mode. For sample results, retain the method and unit together when comparing NTU and FNU readings from different instruments. US EPA Method 180.1, sections 2 and 7 · Thermo Scientific AQ4500 manual, vial handling and measurement

Two kits with a 100 NTU vial
Instrument and kit What the manufacturer lists What to confirm before ordering
Hach 2100Q, StablCal kit 2971210 20, 100 and 800 NTU standards with a 10 NTU verification standard. Hach 2100Q StablCal kit 2971210 Selected curve, required points and bottle or sealed vial format.
Hanna HI98703 or HI93414, kit HI98703-11 AMCO-AEPA-1 cuvettes with expected readings of <0.10, 15.0, 100 and 750 NTU. Hanna HI98703-11 turbidity standards Exact instrument, required points and certificate values for the supplied lot.
The Hach 2100Q kit uses StablCal with 10 NTU for verification. Hanna lists AMCO AEPA-1 cuvettes in its HI98703-11 kit.
Two kits that both contain 100 NTU The values are manufacturer kit examples. Confirm the actual calibration curve, instrument variant and instructions.

For a replacement order, put the kit number, standard type, points and intended use on the same line. A quote that says only “100 NTU standard” leaves those decisions open. Ask for the certificate of analysis and confirm that it covers the supplied lot.

Handle each vial according to its instructions

Read the instructions for the actual pack before mixing. A sealed suspension, a bottled standard and a solid check vial can require different handling. Where the instructions specify resuspension, use the prescribed motion and allow bubbles to clear as directed. Leave sealed vials sealed. If your procedure calls for preparing a working dilution from stock formazin, use the specified water, preparation steps and holding time. Follow the applicable safety data sheet for the material. US EPA Method 180.1, sections 2 and 7 · Hach guidance on Gelex verification standards

Check the glass before blaming the standard. Fingerprints, condensation, scratches and particles on the vial can affect the optical reading. Wipe the outside with a clean lint-free cloth and handle the vial away from the light path. Use the specified fill and orientation. If the manual calls for silicone oil, follow its application instructions. The Thermo Scientific AQ4500 manual gives model-specific guidance on vial handling and indexing. Thermo Scientific AQ4500 manual, vial handling and measurement

Record the vial ID and orientation with an unexpected result. If rotating a vial changes the reading, check the glass and indexing under the manual before using the result to adjust the instrument.

Verify the calibration before accepting sample results

Calibration sets the instrument response using the prescribed standards. Verification checks the response against an assigned value or established baseline. The Hach 2100Q kit lists a dedicated 10 NTU verification standard. Hach describes Gelex as a check material used with an instrument-specific baseline established after calibration. Follow the instructions for the actual check material. Hach 2100Q StablCal kit 2971210 · Hach guidance on Gelex verification standards

Calibrate with the specified points, verify with the method’s check material, then record and review the result.
Keep calibration and verification separate Use the acceptance criterion and review procedure for the method in use.

Suppose a check standard has an assigned value of 10.0 NTU and the meter reads 10.6 NTU. The difference is +0.6 NTU, or +6% relative to that assigned value. This is an arithmetic example. Compare the unrounded result where available with your method’s criterion, using its required calculation and rounding. Record the decision and the criterion together.

If a check falls outside the criterion, retain the original reading. Confirm the vial, lot, expiry, preparation, orientation and instrument mode. Repeat only as the procedure allows and keep the repeat with the original result. If the check still fails, follow your laboratory’s procedure for instrument status and review of affected sample results.

Keep a record the next analyst can use

Write down the calibration and verification materials separately. This helps the next analyst distinguish a new calibration from a repeat check and identify whether a replacement kit changed the vial format, standard material or assigned value.

What to keep with the result
Record field Evidence to retain
Instrument and method Model, serial number, method, mode and reporting unit
Calibration material Kit or part number, material, calibration points, lot and expiry
Vial preparation Pack format, mixing or settling step, condition and orientation
Calibration record Date, analyst, selected curve and stored calibration result
Verification material Check ID, lot and assigned value or documented baseline
Observed check Original reading, difference, acceptance criterion and decision
Follow-up Investigation, repeat result, reviewer and instrument status

Questions from the bench

Can I substitute a polymer vial with the same NTU value?

Confirm the exact standard is supported by the instrument and method. Compare the kit number, suspension, vial format and intended role.

Is stabilised formazin the same as a microsphere standard?

Stabilised formazin is a formazin formulation. Microsphere standards use a different suspension.

Can the verification vial be used to calibrate?

Use each vial for the role specified by the procedure. Hach, for example, designates the 10 NTU vial in the 2100Q kit for verification.

What if a check changes when I rotate the vial?

Inspect the vial and repeat in the documented orientation as your procedure permits. Keep both readings and check the instructions on indexing and vial replacement.